Model: RO Lab / RO Lab Plus
Flow Procedure: PF+AC+RO+AC
Application: Ware washing, Agricultural expe, General biological expe, Aquatic Products feeding, Inlet Water for Unltra pure water machine, Inlet water for sterilizer/ T&H chamber
Purification System: Pretreatment unit Pre-filter (optional) + Special spun fiber filter x1 + Special active carbon block filter x1 + Special active carbon block filter x1
RO Unit: 100 GPD RO Membrane(RO Lab) 2 x 100 GPD RO Membrane(RO Lab Plus)
Subsequent Unit: Post Active Carbon Filter x 1
Pure Water Quality Desalination Rate: 96% – 98%*
TDS: 5 – 10 ppm
Pure Water Outlet: 1: RO Water
Control System: Automatic electronic pressure sensor controlling, RO membrane auto flushing, automatic stop without water, automatic stop when water tank full, automatic cutting off water when pump stops, guaranteeing 24 hr work
Inlet water requirement Tap water: TDS <200ppm, 5-45°C, 1.0-4.0kg/cm²
Output: (25°C): 15 l/hr(RO Lab)* 30 l/hr(RO Lab Plus)*
Instantaneous Output: 1.5 l/min (with pressure tank)
Power: 220V/50Hz, 48W plus model:72W
External Dimension/ Weight H x W x D : 42 x 41 x 22cm / 12-14kg